Loading... Please wait...

Our Newsletter


Olympus EPOCH 1000 SERIES Ultrasonic Flaw Detector

  • Olympus EPOCH 1000 SERIES Ultrasonic Flaw Detector
  • Olympus EPOCH 1000 SERIES Ultrasonic Flaw Detector
  • Olympus EPOCH 1000 SERIES Ultrasonic Flaw Detector
Price:
$9,290.00
Brand:
Condition:
New
Availability:
In Stock
Shipping:
Free Shipping
Minimum Purchase:
1 unit(s)
Quantity:
Bookmark and Share


Product Description

Olympus EPOCH 1000 SERIES Ultrasonic Flaw Detector

The EPOCH 1000 Series Digital Ultrasonic Flaw Detectors combines the highest level of performance for conventional portable flaw detection with the power of phased array imaging. The EPOCH 1000, 1000 iR and 1000 i feature a new horizontal case style with full VGA display, knob and navigation arrows for parameter adjustment, and full EN12668‑1 compliance. The advanced conventional ultrasonic functionality of the EPOCH 1000 series is augmented in the EPOCH 1000 i with phased array imaging capabilities.

Features

- Available with Phased Array Imaging Package
- EN12668-1 compliant
- 37 digital receiver filter selections
- 6 kHz pulse repetition rate for high speed scanning
- Encoded or time-based C-scan option
- Automatic phased array probe recognition
- Intuitive wedge delay and sensitivity calibration for all focal laws
- Programmable analog/alarm outputs
- Designed for IP66 environmental rating
- Horizontal design with navigation panel and knob parameter adjustment
- Digital high dynamic range receiver
- Full VGA sunlight readable display
- ClearWave™ visual enhancement package for conventional A-scan interpretation
- SureView™ visualization feature
- Reference and measurement cursors
- Standard dynamic DAC/TVG
- Standard onboard DGS/AVG

EPOCH 1000 iR
Advanced UT + Phased Array ReadyThe EPOCH 1000 iR provides the same ultrasonic flaw detection capabilities as the EPOCH 1000 with the benefit of upgrading to phased array with simple field-remote activation.

EPOCH 1000 i
Advanced UT + Phased Array Built-inThe EPOCH 1000 i comes standard with the same advanced ultrasonic flaw detection capabilities as the EPOCH 1000, with the addition of a built-in phased array imaging package.

Industries and Applications

- General weld inspeciton
- Crack detection and sizing
- Power generation inspections
- AWS D1.1/D1.5 weld inspection
- Casting and forging defect inspections
- DGS/AVG defect sizing
- In-line inspections
- Composite delamination and defect inspections
- Aerospace and maintenance
- Automotive and transportation


Accessories

The EPOCH 1000 series offers various accessories to allow full functionality of available features.

  • EPXT-C-VGA-6 (U8779019): VGA output cable
  • EP4P/C-USB-6 (U8840084): USB client cable
  • EP1000-C-RS232-6 (U8779197): RS-232 communications cable
  • EP1000-C-26OUT-6 (U8779018): Hardware I/O cable (alarm outputs triggering)
  • EP1000-C-9OUT-6 (U8779017): Hardware I/O cable (analog output)
  • EP4/CH (U8140055): Chest harness
  • EPXT-EC-X: External battery charger
  • EPXT-BAT-L (U8760021): Extra battery (lithium-ion)
  • EP1000-TC-S (U8764078): Small Transport Case
  • EP1000-TC-L (U864079): Large Transport Case



Specifications


GeneralOverall Dimensions (W x H x D)  252 mm x 177 mm x 107 mm;
9.92 in. x 6.97 in. x 4.2 in.
 Weight  3.67 kg (8.1 lb), including lithium-ion battery
 Keypad  English, International, Japanese, Chinese
 Languages  English, Spanish, French, German, Japanese, Chinese, Russian, Italian, Polish
 Transducer Connections  BNC or Number 1 LEMO®
 Data Storage  Onboard up to 10,000 IDs with waveform, standard 4 GB compact flash card (removable)
 Battery Type  Single lithium-ion rechargeable standard
 Battery Life  8 hours (conventional UT mode); 7 hours (phased array mode)
 Power Requirements  AC Mains: 100 VAC to 120 VAC, 200 VAC to 240 VAC, 50 Hz to 60 Hz
 Standby Mode  Adjustable from 15 minutes to 120 minutes with 50% less battery consumption.
 Display Type  Full VGA (640 x 480 pixels) Transflective Color LCD, 60 Hz update rate
 Display Dimensions (W x H, Diag.)  132.5 mm x 99.4 mm, 165.1 mm (5.2 in. x 3.9 in., 6.5 in.)
Inputs/OutputsUSB Ports  1 USB Client, 3 USB Host ports
 RS-232  Yes
 Video Output  VGA output standard
 Analog Output  4 analog outputs, Selectable 1V/10V Full Scale, 4 mA max
 Alarm Output  6 alarm outputs, 5V TTL, 10 mA
 Trigger I/O  Trigger input 5V TTL; trigger output, 5 V TTL, 10 mA
 Encoder Inputs  2-axis encoder line (quadrature)
Environmental RatingsIP Rating  Designed to meet requirements of IP66
 Explosive Atmosphere  Tested to MIL-STD-810F, Method 511.4, Procedure 1. Atmosphere defined per NFPA 70, Article 500 as Class I, Division 2, Group D.
 Shock Tested  IEC 600689-2-27, 60 g's, 6 µsec Half-Sine, 18 Axes total
 Vibration Tested  Sine vibration, IEC 60068-2-6, 5 Hz to 150 Hz @ 0.03 in. or 2 g's Displacement Amplitude, 20 sweep cycles
 Operating Temperature  -10° C to 50° C (-14° F to 122° F)
 Battery Storage Temperature  -20° C to 60° C (-4° F to 140° F)
 Battery Recharge Temperature  0° C to 40° C (32° F to 104° F)
Epoch 1000i Phased Array SpecificationsFocal Laws  61
 Physical Probe  64 elements
 Virtual Probe  16 elements
 Video Filtering  Off, Low, High
 Display Modes  A-scan, S-scan, Linear scan, C-scan, A-scan plus image
 Image Update Rate  60 Hz update for all A-scans; 20 Hz update for all images
PulserEPOCH® 1000, EPOCH 1000iR, EPOCH 1000i (conventional UT mode)Pulser Type Tunable Square Wave
  PRF 5 Hz to 6000 Hz in 5 Hz increments
  Energy Settings 50 V to 475 V in 25 V increments
  Pulse Width Adjustable from 45 ns to 5,000 ns (0.1 MHz) with PerfectSquare™ Technology
  Damping 50, 100, 200, 400 Ω
  Pulser Delay Not applicable
 EPOCH 1000i (PA mode)Pulser Type Tunable Square Wave
  PRF Manually adjustable. Maximum 1520 Hz
  Energy Settings 40 V or 80 V
  Pulse Width Adjustable from 45 ns to 1,000 ns (0.5 MHz) with PerfectSquare™ Technology
  Damping Not applicable
  Pulser Delay 0 to 10 µs, 2.5 ns resolution
ReceiverEPOCH® 1000, EPOCH 1000iR, EPOCH 1000i (conventional UT mode)Gain 0 to 110 dB
  Maximum Input Signal 20 Vp-p
  Receiver Input Impedance 400 Ω ± 5%
  Receiver Bandwidth 0.2 to 26.5 MHz @ -3 dB
  Receiver Delay NA
  Digital Filter Settings Standard filter set (EN12668-1 Test & Compliant): 7 filters
Advanced filter set (not tested to EN12668-1): 30 filters
  Rectification Full wave, positive half wave, negative half wave, RF
  Reject 0 to 80% FSH with visual warning
  Amplitude Measurement 0% to 110% full screen height with 0.25% resolution
  Measurement Rate Equivalent to PRF in all models
 EPOCH 1000i (PA mode)Gain 0 to 80 dB
  Maximum Input Signal 250 mVp-p per channel
  Receiver Input Impedance 50 Ω ± 10%
  Receiver Bandwidth 0.5 to 12.5 MHz @ -3 dB
  Receiver Delay 0 to 10 µs, 2.5 ns resolution
  Digital Filter Settings 6 filters
  Rectification Full wave, positive half wave, negative half wave, RF
  Reject 0 to 80% FSH with visual warning
  Amplitude Measurement 0% to 110% full screen height with 0.25% resolution
  Measurement Rate Equivalent to PRF in all models
CalibrationEPOCH® 1000, EPOCH 1000iR, EPOCH 1000i (conventional UT mode)Automated Calibration • Velocity, Zero Offset
• Straight Beam (first back wall or echo-to-echo)
• Angle Beam (Soundpath or Depth)
  Test Modes Pulse Echo, Dual, or Through Transmission
  Units Millimeters, inches, or microseconds
  Range 3.33 mm to 26,806 mm (0.131 in. to 1054.1 in.) at 5900 m/s (0.2320 in./µs)
  Velocity 635 m/s to 15,240 m/s (0.0250 in./µs to 0.6000 in./µs)
  Zero Offset 0 µs to 750 µs
  Display Delay -59 mm to 25,400 mm (-2.323 in. to 1000 in.)
  Refracted Angle 0° to 85° in 0.1° increments
 EPOCH 1000i (PA mode)Automated Calibration • Velocity, Zero Offset, Sensitivity
• Soundpath or Depth (Zero Offset)
  Test Modes Pulse Echo
  Units Millimeters, inches, or microseconds
  Range 61 focal laws, 3.33 mm to 390.95 mm (0.131 in. to 15.373 in.) at 5900 m/s (0.2320 in./µs)
  Velocity 635 m/s to 15,240 m/s (0.0250 in./µs to 0.6000 in./µs)
  Zero Offset Automatic for all focal laws through calibration, or Manually adjustable from 0 to 100.00 µs (Global Delay Offset mode)
  Display Delay 0 to max range
  Refracted Angle 61 angular focal laws, 0.5°, 1.0°, 1.5°, or 2.0° increments Adjustable from -80° to +80°
GatesEPOCH® 1000, EPOCH 1000iR, EPOCH 1000i (conventional UT mode)Measurement Gates Two fully independent gates for amplitude and time-of-flight measurements
  Measurement Mode Soundpath
  Interface Gate Optional, with Gate 1 and Gate 2 tracking
  Gate Start Variable over entire displayed range
  Gate Width Variable from gate start to end of displayed range
  Gate Height Variable from 2% to 95% full screen height
  Alarms • Positive and negative threshold
• Minimum depth
  Reference Cursors 2 reference cursors for A-scans
 EPOCH 1000i (PA mode)Measurement Gates Two fully independent gates for amplitude and time-of-flight measurements
  Measurement Mode Soundpath, Depth
  Interface Gate Not applicable
  Gate Start Variable over entire displayed range
  Gate Width Variable from gate start to end of displayed range
  Gate Height Variable from 2% to 95% full screen height
  Alarms • Positive and negative threshold (for selected focal law)
• Minimum depth (for selected focal law)
  Reference Cursors 2 reference cursors for A-scans; 4 reference cursors for images
MeasurementsEPOCH® 1000, EPOCH 1000iR, EPOCH 1000i (conventional UT mode)Displayed Measurement 6 locations available (manual or auto selection)
  Gate 1 Thickness, soundpath, projection, depth, amplitude, time-of-flight, min/max. depth, min/max. amplitude
  Gate 2 Same as Gate 1
  IF Gate (optional) Thickness
  Echo-to-Echo Standard. Choose between Gate2-1, Gate2-IF, Gate1-IF
  Other Measurements Overshoot (dB) value for DGS/AVG, ERS(equivalent reflector size) for DGS/AVG, AWS D1.1/D1.5 rating (D), reject value
  DAC/TVG Standard
  DAC Points Up to 50 points, 110 dB dynamic range
  Special DAC Modes 20% to 80% DAC, Custom DAC (up to 6 curves)
  TVG Table Up to 50 points, 110 dB dynamic range, compatible with IF Gate at all PRF settings
  Curved Surface Correction Standard. Tube or bar OD correction for angle beam measurements
 EPOCH 1000i (PA mode)Displayed Measurement 6 locations available (manual or auto selection)
  Gate 1 Thickness, soundpath, projection, depth, amplitude, time-of-flight, min/max. depth, min/max. amplitude
  Gate 2 Same as Gate 1
  IF Gate (optional) Not Applicable
  Echo-to-Echo Standard. Choose between Gate2-1, Gate2-IF, Gate1-IF
  Other Measurements Overshoot (dB) value for DGS/AVG, ERS(equivalent reflector size) for DGS/AVG, AWS D1.1/D1.5 rating (D), reject value
  DAC/TVG Standard
  DAC Points Up to 20 points, 40 dB dynamic
  Special DAC Modes Not Applicable
  TVG Table Up to 20 points, 40 dB dynamic
  Curved Surface Correction

Standard. Tube or bar OD correction for angle beam measurements


Warranty Information

2 Year Warranty